Installation
000000-2227-967-GA-GB-281021
24
Patient environment
The patient environment is the space where direct contact of the patient or
operator with third persons is possible.
When using a further device type in the patient environment, use a device
that complies with IEC 60601-1. If devices are used which do not comply
with IEC 60601-1, an isolating transformer is required to power the device
or connect the devices to additional protective grounding.
3
2
1
5
4
6
1
Peripheral equipment (EN XXXXX and IEC XXXXX)
2
Peripheral equipment (EN XXXXX and IEC XXXXX)
3
ZEISS VISUREF 150 device
4
Patient environment (represented by dotted line, within a radius of 1.5 meter)
5
Power cable with protective grounding
6
Protective grounding
Fig. 7
Patient environment
Radius of 1.5