Operation of the device
000000-2227-967-GA-GB-281021
99
Parameter
Description
Selection and input options
LAN
Page 1/5 Network configuration
Mode
Selection of configuration mode
Manual
DHCP*
IP Address
“Manual” configuration mode: entry of
network parameters
“DHCP” configuration mode: display of data
obtained from the DHCP server
Text input (in “Manual”
configuration mode)
Subnet mask
DNS servers
Gateway
Time server
LAN
Page 2/5 Network folder
Network share
Input of a network share following the syntax
\\<Fileserver>\<Folder>
The file server can be entered via its IP address
or its network designation.
Text input
Example: \\192.168.99.10\RDx or
\\dfszorg445\RDx
User name
Input of a user name to get access to the file
server.
Depending on the configuration of the
network, it may be necessary to specify the
network instance that manages/grants access
authorization in addition to the user name. In
this case use the syntax <user
name>,domain=<domain>. Here, <domain>
stands for the IP address or the network name
of the domain controller.
Text input
Example:
guest1, domain=server1
Password
Input of a user name and the corresponding
password to get access to the file server.
Input path
Path specification for an input folder (for
example, for software update via the network)
Text input
Default: device\input*
Timer
Definition of a time constant for the periodic
reading from the input folder
Selecting the timer
Default: 5s*
Output path 1
Definition of an output path 1 for storage of
measurement data on a network folder. Please
note, that ZEISS VISUREF 150 does not create
any folders. The folders should already have
been created in the network.
Text input
Default: data\txt*
Output format 1
Definition of an output path format for storage
on output path 1 of the network folder
v1.6*
v1.7
XML JOIA
PDF simple
PDF enhanced
Other formats depend on
installation
Output path 2
Definition of output path 2 for storage of
measurement data in a network folder.
Please note that ZEISS VISUREF 150 does not
create any folders. The folders should already
have been created in the network.
Text input of output path
Default*: data\xml