5 Commissioning and First Operating Steps | 5.5 Modifying Gun Parameters
Analytic Gun Mode
Analytic gun mode is suitable for higher probe currents. Overall, the probe current in Analytic gun
mode is about twice the probe current in Normal gun mode.
Info
After switching the gun mode, you can immediately work with the selected gun mode. For ap-
plications, which require a high probe current stability, wait 24 hours until a stability of 0.2 %/
h is reached. It is recommended not to change the gun mode during quantitative specimen
analysis.
Procedure
1. In the SEM Controls panel, select the
Gun
tab.
2. To switch to Imaging mode, click
Imaging
.
3. To return to Analytic mode, click
Analytic
.
5.5.2 Measuring the Probe Current
The probe current is the current that flows through the specimen. It corresponds to the total
number of electrons that hit the specimen.
Measuring the probe current using the Faraday cup ensures that the current displayed in the soft-
ware equals the incident probe current. The Faraday cup consists of a strongly absorbing material
with a cavity covered by a small aperture. If the beam is focused in this cavity, no secondary elec-
trons and no backscattered electrons leave the Faraday cup.
Parts and Tools
Designation
Part no.
Faraday cup
348342-8055-000
Procedure
1. Load the Faraday cup into the specimen chamber. Refer to
2. Evacuate the specimen chamber.
3. Switch on the gun. Refer to
4. Switch on the EHT. Refer to
.
5. Set a magnification that allows transmission of the complete electron beam into the cavity
of the Faraday cup through the aperture of the Faraday cup.
6. From the
Panel Configuration Bar
, select
Specimen Current Monitor
.
à
The
Specimen Current Monitor
window is displayed.
7. Activate the
SCM On
checkbox.
8. Activate the
Spot
checkbox.
à
Green crosshairs are displayed on the image. The crosshairs indicate the position of the
beam spot.
9. Grab the crosshairs and move them into the hole of the Faraday cup.
à
The probe current is measured continuously.
à
The measured probe current is displayed in the
Specimen I
readout.
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006