5 Commissioning and First Operating Steps | 5.2 Switching On the Microscope System
ZEISS
2. Make sure the circuit breakers (F10, F11) are in
the upper position.
3. If the EMO circuit is not installed, plug the CEE
connector into the CEE FEMALE RECEPTACLE of
the mains supply.
4. If the EMO circuit is installed, pull the
EMO
button to release it.
5. Turn the
Main
switch counter-clockwise to the
Reset
position and then back via
OFF
to the
ON
position.
6. Press the green
Start
button.
5.2.2 Starting the Microscope
Info
If the system was powered off for a longer time period, the ion getter pump might fail to start.
In this case, bake-out the gun head (supervisor user rights required) or contact your local ser-
vice center. Refer to
80
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006