
ZEISS
9 Technical Data and Conformity | 9.1 Performance Data and Specifications | Sigma 300 and Sigma 300 VP
9 Technical Data and Conformity
This chapter contains important technical data as well as information on the conformity.
9.1 Performance Data and Specifications | Sigma 300 and Sigma 300 VP
The Microscope System must only be operated in closed rooms. It is recommended to install the
Microscope System in a dark room where artificial illumination, sunlight or other light sources
cannot interfere with image acquisition. The Microscope System should not be installed near win-
dows with direct sunlight or radiators. Compliance with the installation requirements of the Mi-
croscope System and the availability of the requested supplies is the responsibility of the customer
and has to be provided at the time of installation. Due to continuous development, we reserve the
right to change specifications without notice.
The Microscope System must be plugged into a properly installed power socket with protective
earth contact using the supplied mains cable. The protective earth connection must not be im-
paired by the use of extension cables.
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes
Main Components
Length
(mm)
Width (mm)
Height
(mm)
Weight (kg)
column
822
960
1757
636
1133
1076
826
85
Static damping block
180
180
160
44
Quiet mode option
324
324
300
17.0
Pre-vacuum pump
430
250
290
26
Electron Optics
Parameter
Description
SEM resolution
Gemini column in high resolution configuration (max. probe current
20 nA) at optimum working distance:
§
1.0 nm at 30 kV (STEM mode)
§
1.0 nm at 15 kV
§
1.6 nm at 1 kV
Acceleration volt-
age
Range:
20 V to 30 kV
Adjustment:
Continuously variable in 10 Volt steps
Probe current
20 nA high resolution configuration:
3 pA to 20 nA
100 nA high current configuration:
6 pA to 100 nA
Probe current selection
via seven apertures (six apertures in case of
100 nA configuration) with electromagnetic selection and precise
alignment
Optional OptiProbe
module for continuous probe current adjust-
ment
Magnification
Range:
10x – 2,000,000x referenced to Polaroid 5" × 4.5" image for-
mat
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006
147