5 Commissioning and First Operating Steps | 5.9 Shutting down the System
ZEISS
18. When you are finished acquiring the Raman spectra, then move the stage back to the posi-
tion x = 0, y = 0, and z = 0.
à
INFO:
The position x = 0, y = 0, and z = 0 is
the reference position for the correlating
SEM position.
19. To go back to SEM imaging, click
Move to SEM
in the
Raman Correlation
tab of the
SEM Raman Correlation
window.
5.9 Shutting down the System
5.9.1 Finishing the Work Session
To finish your work session, you need to switch off the EHT.
Change to
STANDBY
mode when the microscope is not operated, even for longer periods. The
filament will continue to be heated, and the vacuum in the electron optical column and in the
specimen chamber will be maintained.
The
STANDBY
mode is also the recommended mode to store the microscope. In this case, acti-
vate the
Partial Vent on Standby
checkbox in the
Vacuum
tab of the
SEM Controls
panel. This
can help to prevent oil vapors from penetrating into the specimen chamber during the period of
storage.
NOTICE
Schottky field emitter
If the Schottky field emitter is switched on and off too frequently or inappropriately, its lifetime
is reduced.
4
Avoid switching off the gun during the working week.
4
Use Standby mode for the weekend or a break of up to a week.
4
When using the Standby mode, activate the
Partial Vent on Standby
function.
Prerequisite
ü
The microscope is in ON mode.
Procedure
1. Click the
All:
button in the status bar.
2. Select
EHT Off
from the pop-up menu.
3. Only when interrupting work for longer periods between 2 and 7 days: In the
Vacuum
tab
activate the
Partial Vent on Standby
checkbox.
à
This maintains the gun vacuum, and switches off and protects the turbo pump.
4. Exit the SmartSEM user interface. Refer to
Exiting the SmartSEM User Interface [
5. Close all programs and software.
6. Select
Windows start button > Power icon > Shut down
.
7. At the front of the plinth, press the
Standby
button.