Service Instructions
Confidential and Proprietary
SM-30-0030-A02
Order No. -
18 / 27
Version 07, 2015-12-03
Installation:
1)
Thread the cable through the suspension arm
and secure it position using cable ties.
2)
Mount the potentiometer.
3)
Re-attach the knob and the cap.
4)
Slightly press the halves of the arm together
and secure the cable from the illumination
potentiometer in the clip on the upper half of
the arm, if necessary.
5)
Insert pin (
1
) in lock (
2
).
Fig. 29: Insert pin in lock
6)
Re-insert pin (
3
) and tighten setscrew (
4
).
7)
Re-attach protection flap (
5
).
8)
Adjust the activation point of the microswitch.
9)
Mount the top cover and the locking screw of
the suspension arm.
10)
Re-attach cable conduit (
6
).
11)
Insert OPMI®.
12)
Balance the suspension arm.
13)
Connect cable (
14)
Re-attach covers and locking screws
15)
Re-insert cover or camera control unit
Fig. 30: Re-attach protection flap, cable conduit,
OPMI, balance the suspension arm
6.4
Changing the suspension arm
Removal:
1)
Remove OPMI® from the suspension arm.
2)
Remove locking screw (
7
), cover (
8
) and cable
conduit (
9
).
3)
Remove locking screws (
10
) and cover (
11
).
4)
Remove cover (
12
).
5)
Remove cover or camera control unit (
13
).
Fig. 31: Change the suspension arm
2
1
5
4
3
6
13
10
11
12 8 9
7