System Overview
S100 / OPMI pico
Version 7.0
Page 86
G-30-1781-en
HD video control unit connections only
1
Remote socket (HD)
A foot switch (option) can be connected here.
You can configure the foot switch for the integrated HD video camera
(HD ready 720p) as described in Page 152 or for the integrated video
camera live & streaming control unit (full HD 1080p) as described in
Page 191.
2
Connection for the camera head
The system cable connects the camera installed in the surgical microscope
to the video unit.
Switch the video unit off before plugging or unplugging the system cable
into/from the video unit. The camera may be damaged if the cable is
plugged/unplugged while voltage is still being applied.
3
"HDMI" video output
Digital connection for monitors or recorders from the consumer area.
HDMI (High Definition Multimedia Interface) is used to transfer digital
video data.
4
"YPbPr" video output
Analog connection for high resolution monitors or TV sets
(For the YPbPr cable, see Order Information, Page 254).
5
SD video
"Composite" video output (FBAS). Analog connection for video compo-
nents.
6
USB port
Use only USB media that meet at least USB 2.0 specifications.
7
Service interface
Reserved for ZEISS Service personnel.
8
LAN input/output
Network 100 Mbit
9
Video signal output Y/C
Use the Y/C output if only one analog SD monitor is available and no HD
monitor.
No menus and status information (header bar and footer bar) will be dis-
played in the Y/C video signal output. Use an HD video signal output (DVI
oder YPbPr) for operation and configuration.
We recommend the use of high resolution monitors.
10
DVI output
Analog connection for video components.
Summary of Contents for S100/OPMI pico
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...