Operation
S100 / OPMI pico
Version 7.0
Page 132
G-30-1781-en
Typical procedure
Starting work
• During surgical procedures in the facial area, the patients must wear tint-
ed UV protective eyewear (ordering data see Page 251).
All preparations for use have been completed.
• Check the system for proper function. The process is described in chapter
"Operation - Function test before use".
• Turn on the device with the power switch of the suspension system.
• Select the filter suitable for your application in the illumination beam path.
• Adjust the brightness on the suspension system. Start with the lowest
brightness and slowly increase it to the required but not more than the al-
lowed brightness.
• When focusing, make sure to carefully lower the surgical microscope and
observe the working distance from the patient. The working distance ap-
proximately corresponds to the focal length of the object lens used.
• For focusing, move the surgical microscope over the surgical field into an
ergonomic position within the working distance.
• Select the magnification required using the changer button.
• Look through the eyepieces and lower the surgical microscope until the
surgical field comes into focus. This is to achieve focusing. Ensure a suffi-
cient safety distance to the patient.
• Look through the eyepieces of the binocular tube. Adjust the eyepieces in
such a way that you can see both the edge of the visual field and the mi-
croscope image sharply. Also refer to "Adjusting the surgical microscope".
CAUTION
Damage to the eye caused by high light intensity!
During surgical procedures in the facial area, no light must enter the patient's
eye.
• Make sure to protect the patient's eye from light radiation.
CAUTION
Danger of burnings
Patient may be burnt if the microscope is unattended targeted to the patient
with the light source switched on.
• Never leave the device unattended with the light source switched on.
Summary of Contents for S100/OPMI pico
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...