3.3.5
Configuring the Integrated HD IP Camera
•
Press the Menu button on the control panel.
The OSD menu for adjusting the settings will appear.
The following menus are available for configuring the camera:
AWB
The purpose of white balance is to calibrate the camera to the color temperature of the light
source used. The AWB function of the integrated camera supports the
Continuous / Push
button
modes.
In
Continuous
mode the white balance is carried out continuously. In certain specimens this may
result in incorrect colors.
To ensure that the white balance applies to the light source and not the main color of the
specimen, the
Push button
mode can be set. Here the white balance is carried out only once if
the
AWB / OK
button is pressed outside the OSD menu.
Color temp.
This creates a white balance bias towards yellow (left) or blue (right). If AWB (see above) is set to
push button, the effect of this regulator is only visible after a white balance has been carried out
with the AWB button (outside the OSD menu).
Saturation
The camera's images are in sRGB color space with an 8 bit color depth per channel. The color
saturation per RGB value can be adjusted.
Extreme left:
Gray scales
Extreme right:
Maximum color saturation
Gamma
The gamma correction of the camera can be adjusted. A choice of Gamma 1 and 0.45 is available.
Fig. 23