9 Technical Data and Conformity | 9.1 Applicable Standards and Regulations
9 Technical Data and Conformity
This chapter contains important technical data as well as information on the conformity.
9.1 Applicable Standards and Regulations
Observe the generally applicable local and national safety and accident prevention regulations, as
well as the applicable laws and regulations in your country. The microscope has been designed,
manufactured and tested in accordance with the following regulations and guidelines:
§
The microscopes have been designed, produced and tested in compliance with thefollowing
standards:
- EN / IEC 61010-1
- EN / IEC 61010-2-101
- EN / IEC 61326-1
- EN / IEC 61326-2-6
§
The microscopes meet the requirements of the EC Directive 98/79/EC and 2011/65/EU and
carry the CE mark. Radio interference suppression in compliance with EN 55011 Class A.
§
The instruments must be disposed of in compliance with the WEEE Directive 2012/19/EU.
9.2 Performance Data and Specifications
The Microscope must only be operated in closed rooms. It is recommended to install the Micro-
scope in a dark room where artificial illumination, sunlight or other light sources cannot interfere
with image acquisition. The Microscope should not be installed near windows with direct sunlight
or radiators. Compliance with the installation requirements of the Microscope and the availability
of the requested supplies is the responsibility of the customer and has to be provided at the time
of installation. Due to continuous development, we reserve the right to change specifications
without notice.
The Microscope must be plugged into a properly installed power socket with protective earth con-
tact using the supplied mains cable. The protective earth connection must not be impaired by the
use of extension cables.
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Environmental
Requirements
Environmental effects such as acoustic noise, floor vibration and room temperature have a signifi-
cant impact on the performance of the Microscope and the ability to achieve the specified perfor-
mance with particular regard to long term and sensitive workflows as well as equipment safety.
A site survey can be requested from your local ZEISS service representative prior to the installation.
Location
requirements
Installation site
Exclusively inside buildings
Operating range
Max. 2000 m above sea level
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en
97