98
Description
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Connection panel
1
Remote control socket
for triggering an AUX signal, e.g. to switch on/off an external device
operating at max. 24V/0.5A.
2
Connector for switching component
Possibility of connecting a foot control panel, hand control panel or op-
erating chair with an appropriate footswitch.
3
Potential equalization bolt
4
Indicator window for rated voltage
The voltage shown here must correspond to the rated line voltage pro-
vided on the site of installation. You can adjust the sliding switch using
a suitable tool.
Warning!
Please observe the maximum current consumption of two power outlets
(4) and (5). Only connect medical devices approved by us to outlets (4)
and (5). When using other instruments, make sure that safety is guaran-
teed regarding admissible ground leakage currents. The admissible limit
value of the ground leakage current present in the suspension system's
power cord must not exceed 500 µA in compliance with EN60601-1/IEC
60601-1. CSA/US certification in compliance with UL 60601-1 only allows
a maximum ground leakage current of 300 µA.
5
Power outlet
for medical devices with a current consumption of max. 2 A.
Note:
The current of this power outlet is switched on/off using the S2 power
switch (7).
6
Power outlet
for medical devices with a current consumption of max. 5 A.
7
Power inlet
8
S2 power switch
After the stand has been switched on, the green lamp in the switch is
lit.
9
Strain relief device
The strain relief device prevents inadvertent unplugging of the fol-
lowing electrical connections:
–
power cable