G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Operation
173
Acoustic signals
Service display
In the event of an error, e.g. during the system check after power-on of
the suspension system, an error message appears in the control and dis-
play panel: a wrench symbol and an error code (XXX) are displayed and
three successive beeps are heard.
Inform the service department, telling them the error code and the serial
number of the system.
Motorized control of the surgical microscope is then no longer possible.
All microscope functions except recentering of the X-Y coupling can only
be operated manually.
However, you can still release the magnetic brakes using the appropriate
button on the left and right handgrips of the microscope.
When you press the "MANUAL" key, the surgical microscope can no
longer be operated via the foot control panel, the handgrips or the control
and display panel.
You can continue using the illumination system, but the lamp brightness
is automatically adjusted to a fixed setting. You can use the foot control
panel to switch the illumination system on and off.
Three succes-
sive beeps
– Error message during the software check after
power-on of the suspension system.
– Error message in the case of an internal system er-
ror.
One beep
When the focus or zoom position is saved.
One beep
When brightness level 1.0 is reached.
One beep
After power-on of the suspension system.
Intermittent
beep
Error of the illumination system.
MODE
STORE MANUAL
XXX