Preparations
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Strain relief device on S88 floor stand
Note:
You can secure the power plug and the multipoint connector of the
switching component against inadvertent loosening by installing the two
cables in strain relief device (1).
After you have mounted strain relief device (1), the cable must have the
following length:
– 320 mm from the strain relief device up to and including power con-
nector (9).
– 320 mm from the strain relief device up to connector (10) of the foot
control panel or of an operating chair with appropriate footswitch.
•
Form a loop with the cable as shown in (3).
•
Open flap (4).
•
Feed the cable through opening (5).
•
Close flap (6).
•
Tighten the cable until it encloses flap (7).
•
Check the length of the cable.