OPMI Lumera
®
i on floor stand
Overview of devices
Version 9.2
G-30-1720-en
Page 49
Control elements of the widefield eyepieces
1
Eyecup
for adjusting the distance between eyepiece and eye.
2
Diopter setting ring
adjustable from +5 D to -8 D.
3
Diopter scale
for reading the preset prescription value.
Fig. 9: Control elements
of the eyepieces
NOTICE
Widefield eyepieces with magnetic coupling!
When handling eyepieces that have been removed from the tube, please
keep in mind the usual rules for handling magnets:
•
Do not place the eyepiece near instruments which may be magnetizable.
•
Do not place the eyepiece on sensitive devices such as infusion pumps,
pacemakers, measuring devices, or magnetic data media such as floppy
disks, audio and video tapes, or credit cards.
•
Always store the eyepiece in its original packaging when not using it.
1
2
3
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...