OPMI Lumera
®
i on floor stand
Device data
Version 9.2
G-30-1720-en
Page 151
Electromagnetic immunity for all ME equipment and systems
OPMI Lumera i is intended for operation in the electromagnetic environment
specified below. The customer or the user of the OPMI Lumera i device is
responsible for ensuring that it is operated in such an environment.
Immunity tests
IEC 60601 test level
Compliance level
Electrostatic discharge
(ESD) as per IEC 61000-4-2
±8 kV
contact discharge
±8 kV
contact discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air discharge
Fast transient electrical
disturbances/bursts as per
IEC 61000-4-4
±2 kV for
power lines
±2 kV for
power lines
±1 kV for
input/output lines
±1 kV for
input/output lines
Surge voltages/surges as per
IEC 61000-4-5
±0.5 kV, ±1 kV
phase-neutral voltage
±0.5 kV, ±1 kV
phase-neutral voltage
±0.5 kV, ±1 kV, ±2 kV
phase-neutral to ground voltage
±0.5 kV, ±1 kV, ±2 kV
phase-neutral to ground voltage
Power frequency (50/60 Hz)
magnetic field as per
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Voltage dips, short interruptions,
and supply voltage variations as
per IEC 61000-4-11
0 % UT for 1/2 cycle
at 0°, 45°, 90°, 135°, 180°, 225°,
270°, and 315°
0 % UT for 1/2 cycle
at 0°, 45°, 90°, 135°, 180°, 225°,
270°, and 315°
0 % UT for 1 cycle
and
70 % UT for 25 cycles
Single-phase: at 0 degrees
0 % UT for 1 cycle
and
70 % UT for 25 cycles
Single-phase: at 0 degrees
0 % UT for 250 cycles
0 % UT for 250 cycles
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...