9 Maintenance
OPMI LUMERA 700
9.3 Performing safety inspections
G-30-1673-en - 14.0 - 2018-07-12
181 / 236
9.2.3 Every two years
Component
Activity
Light guide
u
Check to make sure:
• The light guide is intact.
• The light field is uniform.
Light source
Check to make sure the light source
filters are intact.
9.2.4 Every four years
Component
Activity
Support system and
microscope suspension
mount
u
Check to make sure the following
components are free of abrasion
and damage:
• Supporting structures
• Interfaces
• Couplings
• Connection and fastenings of the
spring assemblies
Cables, plugs and switches
u
Check the following functions:
• The insulation and plug connections
of the cables and the device are
intact.
• The cable routing and cable
fastenings are tight.
Electronics:
u
Replace the internal batteries.
Casters
u
Replace the casters.
Light guide
u
Replace the light guide.
9.2.5 Every six years
Component
Activity
Integrated 22" monitor
u
Replace the support arm, including
the gas spring.
9.3 Performing safety inspections
The safety inspection serves to determine and evaluate device
safety. The user of this device is obligated to perform and
document safety inspections in compliance with IEC 62353.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...