7 Operation
Instructions for Use
OPMI LUMERA 700
7.5 Daily operation
160 / 236
G-30-1673-en - 14.0 - 2018-07-12
4. Use the arrow buttons to position the XY coupling.
ð
The XY coupling will immediately move in the corre-
sponding direction. If the video image on the monitor
moves in a direction other than indicated on the button,
the external video camera needs to be positioned differ-
ently.
5. To temporarily save the changes: tap the
button.
7.5.6 Setting the integrated keratoscope ring
The integrated keratoscope ring provides intraoperative visualiza-
tions of corneal curvatures.
Prerequisite
þ
The device includes an integrated keratoscope ring.
Action
1. Open the main menu.
2. Tap the [Light] menu button.
3. Tap the [Keratoscope] submenu button.
ð
The "Keratoscope" menu will appear.
4. To switch on the integrated keratoscope ring: activate the
[Keratoscope On] button.