Instructions for Use
7 Operation
OPMI LUMERA 700
7.1 Safety during operation
G-30-1673-en - 14.0 - 2018-07-12
131 / 236
7 Operation
7.1 Safety during operation
CAUTION!
Risk of injury due to electrical current!
Contact with plug connector contacts may cause electric shock.
u
Never touch the contacts on plug connectors during contact
with the patient.
CAUTION!
Risk of injury caused by high IDIS light intensity!
If IDIS is activated, excessive IDIS light intensity in the eyepiece can
cause glare and may damage the user’s retina.
u
Reduce the IDIS light intensity until an optimum view of the
surgical field is provided.
u
If the IDIS light intensity cannot be reduced, deactivate the IDIS
system.
NOTE
Loss of functionality caused by hardware or software faults!
The XY movement, focus, zoom, light or magnetic brakes may fail,
limiting your ability to operate.
u
Activate manual mode.
u
Contact ZEISS Service.
NOTE
14-function foot control panel failure!
The device cannot be controlled via the 14-function foot control
panel.
u
For wireless 14-function foot control panel: check to make sure
the batteries are sufficiently charged.
u
If the batteries are charged and the foot control panel does not
work: use the cable to operate the 14-function foot control
panel.
u
If the 14-function foot control panel does not work: activate
manual mode.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...