5 Daily startup
Instructions for Use
OPMI LUMERA 700
5.3 Positioning the device in the OR
120 / 236
G-30-1673-en - 14.0 - 2018-07-12
5.3.3 Placing the device in the working position
1
2
Figure 69: Placing the device in the working position
1
Park position
2
Working position
CAUTION!
Risk of injury caused by lowering of the surgical microscope!
If the surgical microscope has not been correctly balanced,
unlocking the magnetic brakes may lead to downward movement,
causing injury to the patient.
u
Only release the magnetic brakes after the system has been
balanced [
111] and the surgical microscope remains
stationary in all positions of the working range.
u
Check the downward travel limit [
113] and maintain suffi-
cient working distance from the patient. The surgical
microscope must not come into contact with the patient even if
it is lowered inadvertently. This applies in particular to installed
accessories such as VISULUX and RESIGHT 700.
u
When unlocking the magnetic brakes, hold the system by the
handgrips to prevent it from unintentionally moving
downward.
Prerequisite
þ
The downward travel limit on the suspension arm is not locked.
Action
1. Release the magnetic brakes.
2. Move the surgical microscope into the desired working
position.
Tip: An unsterile person can use the switching strip to release the
magnetic brakes.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...