64
Preparations for use
G-30-1416-en
OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Connecting the suspension system
•
Check the voltage indicated at (3).
Caution:
The voltage of the stand is factory-set to the rated voltage used in the
country of destination. The rated voltage indicated at window (3) must cor-
respond to the rated voltage available at the site of installation. If this is
not the case, re-adjust the sliding switch using a suitable tool.
Note:
Make sure that the unit has been switched off at the power switch before
plugging in or unplugging any connectors at (1), (4) and (5).
•
Plug the connector of the foot control panel or operating chair into con-
nector (1) of the stand.
•
Connect the stand to line power by plugging the power cord supplied
for this purpose into (6). Only use power outlets which are provided
with a properly connected protective earth contact.
•
If required, connect the stand to the potential equalization contact in
the OR using potential equalization pin (2).