Starting the CMM
6-6
61211-1270302
MMZ M
CMM starting sequence
Proceed in the following order when switching the controller on:
1
Connect the power supply; switch it on with the main switch.
[1]
2
Switch the controller on.
[2]
3
Activate the ram protection.
[3]
4
Turn the drives on.
[4]
5
Select the operating mode.
[5]
6
Switch the computer on and start the measuring software.
Start-up sequence
Connecting the power supply (1)
–
Turn the main switch clockwise to position «1».
The power supply for all electronic components of the CMM is
switched on.
Switching on the controller (2)
1
Turn the rotary switch to the «ON» position.
The indicator lamp of the switch lights up. The controller is on.
2
Wait for approx. 30 s before switching the drives on.
This is necessary because the internal computer of the controller re-
quires a certain amount of time to start. The starting process must
be completed before the drives are switched on. The LEDs on the
control console flash during the starting process.