CT scanner
4-4
61211-8140502
METROTOM 1500 G3 1.4
Rotary table:
Design
Surface-mounted version
Operating position
Vertical
Dimensions
Pallet chuck diameter
[mm]
300
Load capacity
Workpiece weight,
max.
[kg]
50 (incl. workpiece pallet)
Permissible mass mo-
ment of inertia, max.
[kgm²] 0.6
Permissible moment
of tilt, max.
[Nm]
10
Characteristic values
Overvoltage category
III
Pollution degree
2
Protection class
I
Electrical data
Line voltage
400/ 230 VAC (±10%)
Current type
3/N/PE
Frequency
50 to 60 Hz (±3.5%)
Power consumption, max.
6.2 kVA
Typical energy consumption
1.5 - 2.5 kWh
Fuse
C 32A
Environmental conditions
The following conditions must be fulfilled to guarantee flawless opera-
tion of the CT scanner.
Environmental operating temperature
+15 °C to +35 °C
Relative humidity
40% to 70%
Ambient temperature for measuring opera-
tion
20 °C ±2 K
Reference temperature for measurement
18 - 22 °C
Altitude above mean sea level
Max. 2000 m