CHAPTER 3 - ADDITIONAL SOFTWARE TOOLS
LSM 880
SystemMaintenance Tool
ZEISS
10/2014 V_01
000000-2071-464
5
The Maintenance Tool can only be used, if the calibration objective is entered as
APO Calibration LSM
.
If this is not the case, a message will remind the user to select the calibration objective.
2.3
General Operation of the Maintenance Tool
The desired task can be chosen from the list and started with the start button.
Tasks available for LSM
Sharpness Test
Scan Field Test
Scanner Calibration
Pinhole to MBS Calibration
In case a task requires any manual adjustments, the Maintenance Tool guides users to all adjustment
steps.
After the completion of a task, the result is displayed with a traffic light type green/yellow/red indicator.
•
If green is indicated, the task has been fully accepted.
•
If yellow is indicated, repeat the task and try to optimize the settings (e.g. check focus, KÖHLER
illumination, halogen lamps, lasers (no standby) etc.
•
If red is indicated, repeat the test one more time; if it still fails, contact your local LSM service engineer.
Fig. 8
Selection of the calibration objective
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......