LSM 510
Tools
LSM 510 META
Change Filters
Carl Zeiss
03/06
B 45-0021 e
6-19
6.2 Tools
6.2.1 Change
Filters
The
Change Filters
tool is used to update the
filter data in the software after a change of filters
in the reflector turret.
•
Close the LSM 510 software program.
•
Insert the new filter module in the reflector
turret.
•
Double-click on the
Change Filters
icon on the
desktop.
−
The
Emission Filter & Beam Splitter
Control
window appears on the screen. The
name of the currently used database is
displayed in the
System
Database
box, with
the filter type being indicated below for
checking purposes.
−
The
Filter Cubes Stand
panel shows the
Filter-Wheel No.
and the filter positions
available.
−
Use the
Name
and
ID
selection boxes to
enter the filters installed in the individual
positions of the filter wheel.
•
Open the
Name
(or
ID
) selection box of the relevant filter position and select the new filter set from
the list.
•
Click on the
Store
button to accept the new settings.
•
Click on the
Close
button to close the
Emission Filter & Beam Splitter Control
window.
All available filter sets have to registered in the filter list (see
Edit Filter List
function, next page).
Fig. 6-25
Emission Filters & Beam Splitter
Control
window