LSM 510 AND LSM 510 META - SETUP REQUIREMENTS
LSM 510
DuoScan
Carl Zeiss
Space Requirements
LSM 510 META
DuoScan
2-4
B 45-0021 e
03/06
2.1.3
LSM on passive anti vibration table (one microscope, upright or inverted):
2,2 m x 4,2 m
Fig. 2-3
Space requirements for LSM on Passively Damped Anti Vibration Table (system
table with breadboard).
Depending on the ordered table it might be turned by 90°.
(measurements in mm)
The System Electronic Rack contains all electronics for control of the hardware components of the
LSM system, the power supply for the microscope, the scanning unit, the Laser Module V and the
Laser Module RGB equipped with the choice of lasers. The Laser Module RGB is set on top of the System
Electronic Rack. With the extension of LSM
DuoScan
(VIS and/or 405 nm) the Laser Module
LIVE
is
situated in the System Electronic Rack. It substitutes the Laser module V. With the extension of LSM
DuoScan
UV the system has to be set up according to chapter 2.1.5