OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Maintain Menu
LSM 510 META
DuoScan
4-238
B 45-0021 e
03/06
4.10.3.2 Pinhole
Panel
No further software function can be activated and executed during pinhole adjustment.
Pinhole / Description
field:
Selection of pinholes (PH1 to PH4) to be adjusted via the
Pinhole
selection
box, display of the relevant channel in the
Description
field.
Diameter; Pos. X; Y; Z
slider:
Setting of diameter, X-, Y- and Z-position of the pinhole in relation to the
beam path (Z-position can be set only for PH1) using the slider or arrow
buttons, status display for setting procedure: green for ready and red for
busy.
Store current Position
button:
Storage of the current pinhole setting.
Move to stored Position
button:
Pinhole setting is reset to the position last stored.
Adjust Automatically
button:
Automatic pinhole adjustment.
Fast Adjust mode
check
box:
If this check box is activated, the pinhole adjustment is only performed in a
limited area. Used for readjustment.
Fig. 4-238
Pinhole panel