OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-108
B 45-0021 e
03/06
If the entered number of cycles has been processed without a trigger impulse having been given
to end the procedure, the Time Series is finished.
If a trigger signal is given before the cycles have been processed, the Time Series will only be
interrupted.
Waiting for Trigger
will be displayed in the status line. The Time Series can now
be continued via a new trigger signal or ended via
Stop
.
(2)
Stop via Time
To end the Time Series via the time set on the PC
(
Time
button activated), the end time must first be
entered in the
Time
input box.
•
Click on the
Time
input box to open it.
•
Enter the end time via the keyboard. Then click
outside the input box once to close the box.
The Time Series is interrupted when the end time has been reached.
If the entered
Number
of cycles has been processed, the Time Series is finished.
If the number of cycles has not yet been processed, the Time Series is only interrupted.
Waiting
for Start Time
is displayed in the status line. The Time Series can now be continued by entering
a new start time, or finished via
Stop
.
The end time for the Time Series can be changed online.
4.5.6.4
Cycle Delay / Time Interval Panels
Depending on the settings in the
Time
Series
tab
(see
Options
menu,
Settings
), the time series
interval is defined either as a
Cycle Delay
or
Time
Interval
. Accordingly, either the
Cycle Delay
panel or the
Time Interval
panel is displayed in
the
Time Series Control
window.
Cycle Delay
is the interval between the end of
one scan process and the beginning of the next.
Time Interval
is the interval between the
beginning of one scan process and the beginning
of the next.
The
Time Delay
(or
Time Interval)
panel permits
the intervals to be activated and changed.
Fig. 4-89
Stop Series panel
Fig. 4-90
Cycle Delay panel
Fig. 4-91
Time Interval panel