LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-95
If no sufficient coincidence of the two lines can be achieved by the reduction of the scan speed, you have
to calibrate the scanner position signal.
•
Click on the
Maintain
button in the
Main
menu and then on the
Spline
button.
−
The
Calibrate Spline Scan
window is
opened.
•
Bring the generated contour and the scanner
line to coincidence by varying the amplitude or
offset values for X and Y.
•
If necessary, match the free shape curve to the
scanner line.
•
Then click on the
Single
or
Cont.
button to
execute the scan process, with the laser
activated.
A
Line
scan is performed along the defined freehand shape curve, and the intensity profile is displayed at
the bottom of the
Image Display
window.
Irrespective of the length of the defined freehand shape curve, the intensity profile is always calculated
(by interpolation) and displayed in accordance with the pixel value set under
Line Length
(2048 pixels
maximum).
•
Click on the
Stop
button to terminate the scan procedure.
As soon as the free shape curve is modified, the laser excitation is deactivated and the scanner
again starts to trace the newly generated free shape curve.
Fig. 4-74
Calibrate Spline Scan