LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-41
This applies to HeNe and Diode lasers:
•
After selecting the laser, click on the
On
button.
−
The required laser for image acquisition is now available.
After switching on the lasers in the laser control window and their status ready the system can
be used for imaging. However, for quantitative imaging it is recommended to let the system
warm up for 2 to 3 hours.
4.5.2
Microscope Control
The
Microscope Control
(
Micro
button) window permits motorized functions (objective and reflector
change, condensor, filter and diaphragm settings) and the illumination mode (transmitted or reflected
light) of the connected microscope to be controlled via the software.
Without any difference to software control, these microscope functions can also be operated directly on
the stand via the relevant controls. In that case, any changes are recorded by the software and displayed
in the relevant windows / panels.
If you are using
Axioskop 2 FS
MOT
as the basic microscope, the microscope functions cannot
be controlled by the software as the instrument is not motorized except to the z drive. There is
no
Micro
button in case of Axioskop 2 FS
MOT
.
4.5.2.1
Open the Microscope Control Window
•
Click on the
Micro
button.
−
This opens the
Microscope Control
window on the screen.
After conclusion of the conventional setting of the connected microscope, the
Microscope Control
window can be closed again.
•
Click on the
Close
button in the
Microscope Control
window.
−
The
Microscope Control
window will be closed.