3.8
Switch Incubation ON and OFF
•
Before you turn
ON
the incubation, all incubation components you wish to use during the imaging
session must be attached to the Lightsheet Z.1 system before starting ZEN software.
•
Set the
"Incubation" switch
on the remote control to the ON position. Make sure that all incubation
components are switched on.
•
Once ZEN is running you can detach and reattach components.
•
To turn
OFF
the incubation, uncheck the corresponding checkboxes in the
Incubator
tool window
(see CHAPTER 4 SYSTEM OPERATIONl) or deactivate the incubation components by setting the
"Incubation" switch
to the OFF position.
Be aware that all incubation components will continue working with the in
Incubator
tool window
specified parameters after ZEN has been shut down or stopped working. This will keep your sample
alive under precisely defined conditions, even if ZEN stops unexpected during an experiment.
3.9
Illumination and Detection Optics
In the Lightsheet Z.1, one or a pair of illumination optics plus one detection optic are always required.
We recommend the following combinations:
Lightsheet Z.1 illumination
optics 5× / 0.1
Lightsheet Z.1 illumination
optics 10× / 0.2
Lightsheet Z.1 detection optics
5× / 0.16
yes
no
Lightsheet Z.1 detection optics
20× / 1.0
no
yes
Lightsheet Z.1 detection optics
40× / 1.0
no
yes
Lightsheet Z.1 detection optics
63× / 1.0
no
yes