SETUP REQUIREMENTS
ZEISS
Power Requirements
Lattice Lightsheet 7
56
000000-2324-960
04/2021 V_01
2.2
Power Requirements
The following components must be connected to the AC mains grid:
−
power supply unit
−
3 point powerstrip (multipoint connector 2 or incubation components power strip)
−
3 point powerstrip (multipoint connector 3 or storage components power strip)
The following components must be inserted in the 6-point power strip (multipoint connector 1 or system
components power strip, see Fig. 36), which itself is connected to the power supply unit:
−
flat screen monitor
−
control computer (system PC)
−
sCMOS camera
−
stage motion controller
−
camera liquid cooling system
The following components must be inserted in the 3-point power strip (multipoint connector 2 or incubation
components power strip, see Fig. 36):
−
gas mixer controller
−
temp module controller
The folowing components must be connected to the 3-point power strip (multipoint connector 3 or storage
components power strip)
−
analysis computer (storage PC)
−
monitor
The following components must be connected to the power supply unit (PSU):
−
laser module (LM)
−
Lattice Lightsheet 7 module
−
6-point power strip (multipoint connector 1 or system components power strip)
The storage and data analysis PC and respective monitor must be separately connected to the AC network
by further power cables.