System Data
KSK 150 FC
Version 7.1
Page 80
G-30-1158-en
Electromagnetic interference resistance for ME equipment
or systems that are not life-sustaining
The device KSK 150 FCis intended to be operated in one of the electromag-
netic environments specified below. The customer or the user of the device
KSK 150 FC is responsible for ensuring that it is operated in such an environ-
ment.
Surges as per
IEC 61000-4-5
± 1 kV phase-to-neutral
voltage
± 2 kV phase/neutral to
ground voltage
± 1 kV phase-to-neutral
voltage
± 2 kV phase/neutral to
ground voltage
The quality of the supply voltage
should be that of a typical business or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
as per
IEC 61000-4-11
< 5% UT
(> 95% dip of UT) for
0.5 cycle
< 5% UT
(> 95% dip of UT) for
0.5 cycle
The quality of the supply voltage
should be that of a typical business or
hospital environment. If the user of
KSK 150 FC requires continued func-
tion even in the event of interruptions
in the power supply, we recommend
to power the device KSK 150 FC from
an uninterruptible power supply or a
battery.
40% UT
(60% dip of UT) for 5 cy-
cles
40% UT
(60% dip of UT) for 5
cycles
70% UT
(30% dip of UT) for 25
cycles
70% UT
(30% dip of UT) for 25
cycles
< 5% UT
(95% dip of UT) for 5s
< 5% UT
(95% dip of UT) for 5s
Power frequency
(50/60Hz) mag-
netic field as per
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels typical of business
and hospital environments.
NOTE: UT is the AC supply voltage prior to application of the test level.
Immunity
tests
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidelines
Summary of Contents for KSK 150 FC
Page 1: ...ZEISS Colposcope KSK 150 FC Instructions for Use G 30 1158 en Version 7 1 2015 06 11 ...
Page 4: ...KSK150FC Version 7 1 Page 4 G 30 1158 en ...
Page 15: ...Version 7 1 G 30 1158 en Page 15 KSK 150 FC Safety Measures ...
Page 17: ...Version 7 1 G 30 1158 en Page 17 KSK 150 FC Safety Measures 1 5 1 1 1 6 2 2 4 2 3 ...
Page 19: ...Version 7 1 G 30 1158 en Page 19 KSK 150 FC Safety Measures 7 10 8 11 12 9 8 ...
Page 21: ...Version 7 1 G 30 1158 en Page 21 KSK 150 FC Safety Measures 19 18 17 13 14 15 16 ...
Page 27: ...Version 7 1 G 30 1158 en Page 27 KSK 150 FC System Overview 1 3 2 1 3 4 2 5 6 7 ...
Page 29: ...Version 7 1 G 30 1158 en Page 29 KSK 150 FC System Overview 3 4 5 6 7 1 2 8 9 10 ...
Page 31: ...Version 7 1 G 30 1158 en Page 31 KSK 150 FC System Overview 1 3 6 2 5 4 8 10 11 9 12 7 ...
Page 37: ...Version 7 1 G 30 1158 en Page 37 KSK 150 FC Preparations for use ...
Page 45: ...Version 7 1 G 30 1158 en Page 45 KSK 150 FC Preparations for use 1 2 ...
Page 63: ...Version 7 1 G 30 1158 en Page 63 KSK 150 FC Care and Maintenance 4 1 5 3 2 ...
Page 66: ...Care and Maintenance KSK 150 FC Version 7 1 Page 66 G 30 1158 en ...
Page 84: ...System Data KSK 150 FC Version 7 1 Page 84 G 30 1158 en ...
Page 90: ...Index KSK 150 FC Version 7 1 Page 90 G 30 1158 en ...
Page 91: ...Version 7 1 G 30 1158 en Page 91 KSK 150 FC Blank page for your notes ...