INTRABEAM Water Phantom
Operation
Operation
Measurement workflow
The most important steps for measuring the depth dose curve and the isot-
ropy of the XRS 4 with the INTRABEAM Water Phantom are:
–
For the emission of X-ray radiation in the Water Phantom, your
INTRABEAM must perform the prescribed verifications on the XRS 4 being
used, as if for a proper treatment. In other words, for the INTRABEAM 600
see "Quality assurance".
–
Fixing the rotary feedthrough:
–
Tighten both fastening screws of the XRS mount evenly by hand.
–
Tighten the clamping screw by hand to fix the rotary feedthrough in
one of its other seven detent positions.
–
Insert the XRS 4 into the XRS mount in the rotary feedthrough.
NOTE
You may obtain incorrect measured values if you do not observe the specified
workflow.
•
Perform the measurements only by means of the outlined workflow.
NOTE
Problems in the workflow due to a doubtful and/or unclear display of settings
and results may lead you to the decision to temporarily shut down your IN-
TRABEAM.
•
Contact ZEISS Service if you have any questions regarding the operation
of the water phantom.
Summary of Contents for INTRABEAM
Page 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Page 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Page 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Page 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Page 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Page 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...