Humphrey Matrix Visual Field Instrument
35
Rev. D 08/21/06 PN 701692
CALIBRATION
The Humphrey Matrix Visual Field Instrument does not require scheduled calibration. The
instrument calibration is automatically checked each time the instrument is powered ON and at
the start of each test to be sure the unit is properly calibrated. If the instrument detects the need
for calibration, the Operator LCD Display will display a needs calibration warning. If not calibrated
when the needs calibration warning is displayed, the unit will continue to operate normally until
the unit reaches the calibration limits. Once the calibration limits are reached, the unit will not
operate normally until a calibration is completed successfully. Perform calibration at any time,
not only when requested by the instrument. Calibration will take approximately 15 minutes to
complete.
To calibrate the instrument,
select
the
Help (F6)
Screen, and then
select
the
Calibration
Button.
Before selecting the
Calibrate
Button in the
Calibration
Screen (see Figure L), make sure the
Patient Eyepiece is covered with the Calibration Cap. Once the Patient Eyepiece is covered,
select
Calibrate.
There is a prompt to ensure the Patient Eyepiece is covered. If the Calibration
Cap is not available, substitute something that will temporarily block light from entering the
Patient Eyepiece (black cloth over Patient Visor, etc.) or perform the calibration in a completely
darkened room. The Operator LCD Display will indicate if there is too much ambient light to
complete the calibration. A
Calibration Progress
indicator is provided.
Figure (L)
Summary of Contents for Humphrey matrix
Page 25: ...Humphrey Matrix Visual Field Instrument 21 Rev D 08 21 06 PN 701692 Figure C...
Page 63: ...Humphrey Matrix Visual Field Instrument 59 Rev D 08 21 06 PN 701692...
Page 64: ...60 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 65: ...Humphrey Matrix Visual Field Instrument 61 Rev D 08 21 06 PN 701692...
Page 66: ...62 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 67: ...Humphrey Matrix Visual Field Instrument 63 Rev D 08 21 06 PN 701692...
Page 68: ...64 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 69: ...Humphrey Matrix Visual Field Instrument 65 Rev D 08 21 06 PN 701692...
Page 70: ...66 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 71: ...Humphrey Matrix Visual Field Instrument 67 Rev D 08 21 06 PN 701692...
Page 72: ...68 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...