Table I.4 General Networking Error Messages
Error Message
Possible Solution
A hard drive database failure has occurred. The hard drive
database must be repaired before backup to file server can be
performed.
• Refer to
“How to Handle Database Failures,” on page 11-10
A hard drive database failure has occurred. The hard drive
database must be repaired before restore from file server can
be performed.
• Refer to
“How to Handle Database Failures,” on page 11-10
A hard drive database failure has occurred. The hard drive
database must be repaired before archiving can be
performed.
• Refer to
“How to Handle Database Failures,” on page 11-10
A hard drive database failure has occurred. The hard drive
database must be repaired before data retrieval can be
performed.
• Refer to
“How to Handle Database Failures,” on page 11-10
A network address must be of the form XXX.XXX.XXX.XXX
where XXX is a number between 0 and 255.
• Re-enter the IP address in the correct format.
Cannot access the remote host. Check network connection
and try again.
• Check all cable connections.
• Verify the server is powered up and running.
• Verify accuracy of the IP address for your server and re-enter the
address, if necessary.
Cannot access the specified folder. Check permissions and try
again.
• Verify that the user has access privileges for the specified folder.
• Change to a folder for which you have access privileges.
Cannot find server. Check network connection and settings
and try again.
• Check all cable connections.
• Verify the server is powered up and running.
• Verify accuracy of the IP address for your server and re-enter the
address, if necessary.
Cannot find specified folder. Ensure folder exists on the server
and try again.
• Verify that the specified folder exists on the network server.
• Ensure that you have not misspelled the folder name.
• Re-enter the folder name or use a different folder.
Cannot login to server. Check user name and password and
try again.
• Re-enter your User Name and Password for the server.
• If problem persists, verify User Name and Password with the
System Administrator.
Cannot write to remote folder. Check permissions on the
server and try again.
• Verify that you have write-access to the specified folder.
• Get needed write-access privileges from System Administrator.
Could not obtain IP address from the DHCP server.
• Check all cable connections.
• Verify the server is powered up and running.
• Verify accuracy of the IP address for your server and re-enter the
address, if necessary.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......