9 Technical Data and Conformity | 9.4 Performance Data and Specifications | GeminiSEM 560
Standard VP
(option)
Parameter
Description
Vacuum range
Vacuum range from 5 Pa to 60 Pa, adjustable in steps of 1 Pa
VP detection
At least one of the following detectors must be selected in the stan-
dard VP option:
§
High speed Variable Pressure Secondary Electron Detector (VPSE)
§
High efficiency Cascade Current Detector (C2D)
§
Backscatter Electron Detector (AsB/aBSD/aBSD-LH)
NanoVP (option)
Parameter
Description
Beamsleeve
Retractable beamsleeve aperture below the objective lens to shorten
the beam gas path length and to enable InLens detection in VP mode.
Vacuum range
Vacuum range from 5 Pa to 500 Pa, adjustable in steps of 1 Pa
VP detection
Additional to the detectors selected in standard VP option, the VP-
BSD detector is available in NanoVP mode.
9.4 Performance Data and Specifications | GeminiSEM 560
The Microscope System must only be operated in closed rooms. It is recommended to install the
Microscope System in a dark room where artificial illumination, sunlight or other light sources
cannot interfere with image acquisition. The Microscope System should not be installed near win-
dows with direct sunlight or radiators. Compliance with the installation requirements of the Mi-
croscope System and the availability of the requested supplies is the responsibility of the customer
and has to be provided at the time of installation. Due to continuous development, we reserve the
right to change specifications without notice.
The Microscope System must be plugged into a properly installed power socket with protective
earth contact using the supplied mains cable. The protective earth connection must not be im-
paired by the use of extension cables.
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes
Main Components
Length
(mm)
Width (mm)
Height
(mm)
Weight (kg)
column
1000*
1540
1866.6
780*
980
980
variable
97.2
Static damper
180
180
160
12.0
Pre-vacuum pump
430
376
216
24.5
Chiller (optional, water- or
air-cooled)**
530
640
740
90
Compressor**
281
139
239
25.0
* Depending on configuration
** Actual measurements are country dependent (different manufacturers).