3 Product and Functional Description | 3.3 Optional Components and Accessories
3.3.2 Control Panel
Purpose
The control panel allows direct access to 14 of the most frequently used functions. It integrates a
full sized keyboard, 11 turning knobs, and 8 push buttons.
Position
The control panel is placed on the work desk.
1
2
3
6
7
8
4
5
11
10
9
Fig. 28: Control panel
1
Stigmator X | Stigmator Y
Shapes the beam roundness by changing the stigmation deflectors.
2
Aperture X | Aperture Y
Adjusts the mid column shift and tilt deflectors for aligning the beam along the column
axis.
3
Scan Rotate
Rotates the scanning pattern 360° continuously.
This turning knob has a push button function to deactivate the scan rotate function and
reset the scan rotation to 0°.
4
Shift X | Shift Y
Shifts the scanned region of the specimen in the X and Y directions.
5
Brightness | Contrast
§
Brightness
Adjusts the image acquisition chain offset for the currently selected detector. Each
configured detector stores its own brightness.
§
Contrast
Adjusts the gain of the currently selected detector.
6
Magnification | Reduced
§
Magnification
Adjusts the magnification of the system.
§
Reduced
Changes the scan field to a reduced area. The size of the area is determined by the
current sub scan area settings.
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000