3 Product and Functional Description | 3.2 Main Components
ZEISS
Fig. 15: Chamber
camera disabled as indicated by a pause sign (e.g. if a diode detector is
selected).
3.2.5 Specimen Stage
Purpose
The motorized eucentric specimen stage is used to navigate the specimen inside the specimen
chamber.
Position
The specimen stage is mounted on the chamber door. If the chamber door is closed, the speci-
men stage is inside of the specimen chamber.
1
2
Fig. 16: Specimen stage with dovetail fitting for precise fitting
1
Specimen holder
2
Specimen stage
Operating
Principle
When moving a tilted stage in X/Y direction, no displacement (moving out of focus) occurs. When
a pre-tilted specimen holder is used, X/Y movements require an adjustment of the Z position to
stay in focus.
The stage can be operated using the
or using the SmartSEM software.
Axis
Description
Movement
X-axis
Movement towards or away from the cham-
ber door (horizontal movement in the image)
Y-axis
Movement
Height
Vertical movement (movement towards or
away from the focal plane of the image)
Height
Movement of the specimen surface into the
tilt axis at various working distances
Rotation
Stage rotation parallel to the X-Y plane
Tilt
Stage tilt about an axis parallel to the X axis
42
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000
Summary of Contents for Crossbeam 550
Page 135: ......