2.4.1.4. Start button S2 (with optional EMO function only)
The
S2
button (
1
) is located at the rear of the plinth
underneath the
MAIN switch
.
It is required to confirm the setting of the
MAIN
switch
.
It cuts off all devices connected to the EMO box
from the mains power supply. The main switch can
be secured against re-activation.
The main switch guarantees an ampere interrupt-
ing capacity (AIC) of at least 10000 A rms.
2.4.1.5. EMO button (optional, but mandatory for compliance with SEMI S2)
The Emergency off button (
1
, EMO button) is
located on the plinth adjacent to the specimen
chamber.
The EMO button is to be pressed in an emergency
to cut off mains power to the microscope and all
devices connected to the EMO box via the AC-
Unit.
The EMO button must always be readily accessi-
ble and operable.
1
1