8.4. Power circuit
8.4.1. Checking the circuit breakers
Reasons
•
Circuit breaker is tripped (lower position)
Procedure:
1
Check the circuit breakers at the back of the plinth.
2
If one of the circuit breakers is tripped, push upwards.
Note
If this does not solve the problem, contact your local Carl Zeiss service for assistance.
IMPORTANT
Tripped circuit breakers may be a hint for an electrical problem in the microscope or with
a electrical component connected to one of the IEC Plugs at the panel of the AC-Unit.
If a circuit breaker is tripping again, switch off the microscope completely and contact the
Carl Zeiss service for assistance.
No.
Value
Circuit
F1
10 A
Power supply unit
F2
10 A
PC, WDX, EDX, AUX 1 -4
F3
10 A
Airlock, pre-vacuum pump
F4
10 A
Internal heaters
Table 8.1: Circuit breakers