Corona process | User Manual
Maintenance
Corona process
7
White calibration
Position the standard holder
2
with the white standard
3
at position
P1
.
Loosen the clamping screw
5
and slide the white standard along its sliding
axis
1
against the measuring window
4
of the
Corona process
until it
touches the measuring window.
Fig. 34
4
3
2
1
5
Positioning the white standard
Fasten the white standard
3
in this position by tightening the clamping
screw
5
.
Fig. 35 Continuing the white calibration
Press the
OK
button to continue the process.
Sliding axis of the white/
black standard
Standard holder
White standard
Measuring window
Clamping screw