Corona process | User Manual
Maintenance
Corona process
7.2 Changing halogen lamps
Warning
ELECTRICAL ENERGY HAZARD
Make sure that the power supply is disconnected – the entire
measuring system must be switched free of voltage.
Protect the measuring system against unintentional activation of the
power supply by third parties.
Caution
GENERAL RISK
All specified activities may be performed only by Service staff of
Carl Zeiss Spectroscopy GmbH or by properly instructed personnel.
Caution
THERMAL ENERGY HAZARD
Let the halogen lamp to be replaced cool off for approx. 10 minutes
before removing it.
Caution
SENSITIVE OPTICAL COMPONENTS
Perform work on optical components only in a very clean working
environment.
Be careful not to touch the glass bulb of the new halogen lamp
without putting on cotton gloves first.