6 Configuration
Instructions for Use
CLARUS 500
6.7 Configure Capture Settings
104 / 144
2660021165005 Rev. A 2017-08
NOTE
Unprotected devices
may be at risk from unauthorized individuals.
u
Select and set the Password strength setting in the User man-
agement screen.
6.7 Configure Capture Settings
Select the
Acquisition
button on the
Specific Settings
tab to con-
figure acquisition settings.
Figure 21: Acquisition Settings
Setting
Explanation
Show last captured
image for (x) seconds
Sets how long (in seconds) an image is dis-
played after capture
Blink Fixation target
for (x) seconds
Sets how long (in seconds) the fixation target
blinks when the operator selects the
Blink
but-
ton on the
Acquire
screen
Autofocus
Turns autofocus on by default
Reset to Factory De-
fault
Returns all acquisition settings to the default
settings
6.8 Configuring Import and Export
Select
Settings
>
Specific Settings
>
Import Export Settings
to
configure import and export settings.
Setting
Explanation
Export to FORUM
If this setting is on, exam data is automatically
exported to FORUM at the configured fre-
quency