Specialized and Integrative Visualization Tools
8-55
cáÖìêÉ=UJPV=eçï=íç=~ééäó=íÜÉ=^ÇÇ=^åÖäÉ=qççä
The Anterior Chamber Angle measurements provided by CIRRUS HD-OCT are not intended
to be a substitute for gonioscopy, which is the current reference standard for evaluating the
anterior chamber angle configuration. For example, during gonioscopy, end users can view
the ACA through a mirror/prism under dynamic conditions, and can examine the angle
over the full extent. CIRRUS HD-OCT provides an optical image of the angle as represented
(and analyzed) by the software at a single location.
In Wide Angle-to-Angle Analysis
In addition to providing a wide view image of the anterior chamber, angle tools and
calipers are provided with the Wide Angle-to-Angle Analysis to allow angle measurement
as shown in
.
Specialized and Integrative Visualization Tools
High Definition Images Analysis
For all raster scans, High Definition Images is the only available analysis.
When you select a high-definition raster scan from the Analysis screen scan list, the High
Definition Images analysis for that scan is automatically displayed.
The High Definition Images analysis screen displays the following:
• Signal strength, scan length, scan angle, line spacing, and image tools
• A fundus image with an overlay showing the location of the scan lines
• For multi-line scans, thumbnails of the scan line images
• A single larger image of the selected scan line for HD 1 Line 20x scans
B
A
C
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...