WARNING – GENERAL HAZARDS 

These quick instructions are intended as an overview. Instructions for the safe operation of the device are to be found in the user's manual.

 

F

 

Please read the CIRRUS photo user manual carefully before using this product. 

 Select acquisition mode and scan parameters

 

 

 

 

 

 

Select this mode for 

OCT imaging only 

Select this mode 

for combined OCT 

+ fundus imaging 

Select this mode 

for fundus imaging 

only. 

 

 

Select the desired imaging mode. 

 

Advanced tip: 

 

For patients who have difficulty fixating, take OCT and fundus images as 
separate acquisitions (

CIRRUS 

and 

Photo

), and combine them using 

 or 

.

 

 

 

Align and fo cus the instrument using working distance dots

 

 

 

 

 

 

 

Pull the instrument back from the patient and place patient‘s chin in the chinrest.
Adjust table height and use eye level marks to adjust height of chinrest.
Use joystick to align instrument with patient pupil (use alignment circles for best 

positioning). 

 

Move instrument towards patient. The focusing aid and working distance dots (two 

white dots) will appear together with the fundus image.
Bring the working distance dots roughly into focus.
If AutoFocus fails, use knob on side of instrument (bars should be aligned 
horizontally).
Bring the working distance dots into focus by turning and moving the joystick and 

align the working distance dots within the horizontal lines.

 

 

Acquire image(s)

 

 

 

 

 

 

 

Instruct patient to blink once and to look directly at the fixation target 

.

 

Click joystick button to capture image(s).

 

Advanced tip:

 

To modify OCT scan pattern location, press arrow (

) on keypad before capturing 

images. Use controls on OCT fundus image to change the scan location and 

rotation/size (HD Raster only).

 

 

 

Grab and drag scan 

pattern to preferred scan 

location.

Grab edge of scan pattern to 

resize the raster scan length 

and spacing. For 1 Line Raster, 
drag spacing to zero.

Press 

<!>

 on keypad or the 

Optimize

 button in display. 

Use the scroll wheel of your 

mouse to center the scan.

 

 

 

Review the captured scan, then save or try again.

 

 

 

 

 

Evaluate the B-scan quality and the signal strength (should be 6 or greater).

 

Check that the OCT B-scans in the display windows appear centered and that no 

data is cut off or missing.

 

Click 

Try Again

to repeat scan acquisition, if necessary.

 

Click 

Save 

to save the OCT scan (fundus images are automatically saved).

 
 

false 

correct 

CIRRUS

 photo Scan Acquisition (Models 600 and 800)

Quick Reference Guide  

C

M

Y

CM

MY

CY

CMY

K