Coordinate measuring machine
3-3
CenterMax Operating Instructions
61211-1020202
Connection data
Electric data
Compressed air
Certain requirements have to be met before connecting the com-
pressed air in order not to impair the function of the CMM.
Travel speed
Line voltage
100, 110, 115, 120, 125, 230,
240 VAC (±10%)
Type of current
1/N/PE
Frequency
50 to 60Hz (±3.5%)
Power consumption
max. 2500VA
Fuse
for 230V or 240V
C 25A
for 100V or 125V
25A
Requirements
Pressure
6 - 10bar
Degree of purity
99%;
without
solid particles, gas-
eous proportions of oil and water
Air consumption
max. 20l/min with 5bar
Creep speed: SLOW function
0 to 5 mm/s
Setup mode: MAN operating mode
0 to 70mm/s
Series measuring operation: AUTO operating mode
Axis
Vector
max. 300mm/s
max. 425mm/s (or 520mm/s)
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......