Special measures
8-5
CenterMax Operating Instructions
61211-1020202
Special measures
After collision
• Eliminate the cause of the collision.
• Switch on the drives.
• Turn the rotary switch on of the control cabinet clockwise and
hold it until the indicator lamp located above lights up.
• Initialize the CMM if the cause cannot be eliminated by means of
the above measure.
➤
Operating instructions for the measuring software.
• Then switch on the drives.
Collision during probe change
A collision occurring during probe change may be caused by missing
intermediate positions.
• Reallocate the storage position to the probe and choose the inter-
mediate positions so that probe change can be carried out without
collision.
Probe head outside the measuring volume
If the probe head moves in the z axis outside of the measuring vol-
ume, the drives are switched off. The drives can only be switched on
again after the probe head has been positioned inside the measuring
volume.
Returning the probe head to the measuring volume
To return the probe head to the measuring volume, proceed as fol-
lows:
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......