7-4
Measuring operation
61211-1020202
CenterMax Operating Instructions
Probing with VAST
The probe must be moved back counter to the probing direction after
each single point probing.
Measurement in static
status
The measuring data are not accepted immediately after probing.
Instead, they are initially polled following a damping period of approx.
one second. I.e. the actual measurement occurs only after the
machine reaches a static state.
Multipoint measurement
Multipoint measurement is only possible with the VAST probe system.
During multipoint measurement, the probe is moved perpendicular to
the probing direction.
The probing axis runs parallel to the workpiece surface. During mea-
surement, the probe element constantly remains in contact with the
workpiece surface. The measuring force remains effective until prob-
ing is completed.
A
Lowering the probe and probing the workpiece; with the left joystick.
B
-
D
Move the probe over the workpiece with the right joystick. The
position in the z direction remains unchanged.
D
Lifting the probe.
A
B
D
A
C
B
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......