4 Installation | 4.4 Assembling Components to the Binocular Tube
ZEISS
4.4 Assembling Components to the Binocular Tube
The following components can be inserted into the tube:
§
eyepieces
§
auxiliary microscope
§
pinhole diaphragm
1
2
3
4
5
1
Dust cap
2
Eyepiece
3
Auxiliary microscope
4
Pinhole diaphragm
5
Dust cap
Procedure
1. Remove both dust caps
1
/
5
from the tube.
2. Remove both eyepieces
2
from the box and insert them into the eyepiece socket of the
tube to the stop.
NOTICE
Before inserting Pol eyepieces with tubes without upright reticles, the
orientation screw on the reverse side of the eyepieces must be unscrewed. The
eyepieces cannot otherwise be fully inserted.
3. Instead of an eyepiece insert an auxiliary microscope
3
or pinhole diaphragm
4
in
one eyepiece socket.
Proceed in the reverse order for removal.
66
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001